Size: |
249 mm (W), 97 mm (D), 235 mm (H)
9.8" (W), 3.8" (D) 9.25" (H)
|
Unit weight: |
2.5 kg (5.5 lbs) (without carry bag) |
Power requirements: |
AC adaptor: 100-240V AC(+/- 10%)
50-60 Hz in, 24V DC, 5.0A out
DC adaptor: 11.5 - 16V DC in, 24V, 5.0A out
(Important: see accessories list for model and part numbers for AC and DC power supply packs.)
|
Purity: |
87% - 96% at all flow rates, over operating conditions |
Setting: |
User adjustable in 0.5 increments from 1.0 to 5.0 |
Inspiratory trigger sensitivity: |
-0.12cm/H20 |
Setting indicator: |
LCD |
Maximum oxygen discharge pressure: |
20.5 psi |
Humidity range: |
5% to 93% ± 2% non-condensing |
Disconnect Device: |
Power supply input connector |
Operating altitude: |
0m to 4,000m (0′ to 13,000′) relative to sea level, 1060 down to 575 mbar |
Sound level: |
Sound Pressure level of 37dBA at setting 2 and 18 BPM: (when measured at 1 m from front of device)
Sound Pressure level of 48.7dBA/Sound Power level of 56.7dBA at setting 5
|
Type of protection (electrical): |
Class II |
Degree of protection (electrical): |
Type BF |
Degree of protection (water): |
IP22 in carry bag (protection against small objects and tilted dripping water)
IP20 out of carry bag (protection against small objects and no protection against water entering the concentrator)
|
Degree of safety (flammable anaesthetic mixture): |
Not suitable for use in the presence of a flammable anaesthetic mixture
|
Operating temperature: |
Continuous operation at temperatures between 5°C (41°F) and 40°C (104°F)
|
Storage temperature: |
Between -20°C (-4°F) and 60°C (140°F)
|
Alarm sound pressure range: |
68.1 to 69 dB(A)
|
Alarm system delays: |
Less than 10 seconds after detection (low oxygen alarms if oxygen is less than 82% volume fraction at specified environmental conditions)
|
Oxygen concentrator status indicator: |
High priority alarm that indicates when oxygen concentration drops below 82%
|