MEDIWET - NOT AVAILABLE
The humidifier for oxygen therapy is a device allowing an increse in the relative humidity in the oxygen supplied to the patient; this gas for medical use normally lacks a sufficient humidity grade to be physiologically tolerated.
The jar is made of polycarbonate, or polysulphone the cap with inlet adaptor of polypropylene and inlet connection of polypropylene + coated brass.
Art. Nr. | Description |
K294401 | Mediwet II 121 ºC inlet connection 9/16 UNF F |
K294416 | Mediwet II 121 ºC inlet connection G 3/8 F |
K294402 | Mediwet II 134 ºC inlet connection 9/16 UNF F |
K294432 | Mediwet II 134 ºC inlet connection G 3/8 F |
TECHNICAL DATA
Contents: | Only sterile water |
Dimensions: | Height 190 × Width 85 (incl hose nipple) × ∅ 60 mm |
Weight: | 150 g |
Capacity: | 200 ml of water |
Material: | Jar polycarbonate (autoclavable at 121 °C) Jar polysulphone (autoclavable at 134 °C) Lid and outlet hose nipple: polypropylene |
Inlet nut: | The brass thread and it is injected to the plastic hand nut |
Diffuser: | Polyethylene |
Hose: | Silicone |
Gasket: | O-ring NBR |
Outlet connection: | Tapered hose nipple for hose 6 × 9 mm (recommended length 2 m) |
Cleaning: | Water, non abrasive detergent. Never use solvents. |
Disinfection: | An alcoholic solution, or other solution compatible with the material according to the disinfectant manufacturer. |
Autoclave: | Polycarbonate version 121°C for 20 minutes Polysulphone version 134°C for 18 minutes |
Special case: | |
Diffuser: | Exchange at every cleaning - do not sterilise! |
Maintenance: | Check that the humidifier is whole and air tight before use. Monthly exchange of gaskets is recommended. Exchange the diffuser when its microperforations no longer exist. |
Durability: | Minimum 20 autoclave cycles undr the condition that all instructions accompanying the product are adhered. |
SPARE PARTS
Art. Nr. | Description |
K301059 | Jar 121 ºC |
K301062 | Jar 134 ºC |
K301078 | Lid connection G3/8 |
K301064 | Lid connection 9/16 |
K294433 | Inlet connection gasket G3/8 (10 pcs) |
K294407 | Inlet connection gasket 9/16 UNF (10 pcs) |
K294408 | Bottle gasket 121 ºC (10 pcs) |
K294409 | Bottle gasket 134 ºC (10 pcs) |
CONSUMABLES
Art. Nr. | Inlet connection |
K294434 | Complete diffuser (tube 13 cm and diffuser, 10 pcs) |