MANUAL AND AUTOMATIC RESUSCITATION SYSTEM

  • MARS II is a leading resuscitator developed for healthcare professionals and first responders. 
  • MARS II is specifically designed to help emergency personnel, respond to patients that require resuscitation. The device can be used in confined spaces, low oxygen/toxic environments.
  • MARS II meets the requirements of ISO 10651-5:2006 and European Resuscitation Council Guidelines for Resuscitation 2010 for ventilatory resuscitators.
  • The MARS II is available in all regional gas-standards.
  • MARS II can be used for children (20 kg), small adults and adults.

FEATURES

  • Simple to use
  • Robust design to withstand harsh environments
  • Can be enabled for automatic or manual (CPR) resuscitation


735100000588 Mars II EN leaflet

TECHNICAL DATA

MARS II CONTROL MODULE

Gas: O2
Material: Brass, aluminium main inside parts, abs cover
Dimensions: 165 × 110 × 63 mm
Weight: 1300 g
Regulator Inlet Connections: According to regional high pressure standards
Inlet pressure
(without reg.):
3.6 - 6 bar at 100lpm
Working pressure: 3 bar
Inlet connection (module): Regional standard low pressure probe
Regulator performance: min 100 lpm and min 3 bar
Inlet filter: 30μm
Time to revert to automatic resuscitation: 5 - 7 seconds
Gas consumption: 9.6 lpm (Adult mode)

MARS II DEMAND VALVE

Material: Polycarbonate, silicone, rubber, stainless steel
Dimensions: 120 × 50 × 70 mm
Weight: 175 g

Inspiratory resistance
at 5 lpm:

 -2.3 hPa (-2.3 cm H2O)
Inspiratory resistance
at 60 lpm:
-4.4 hPa (-4.4 cm H2O)
Expiratory resistance
at 60 lpm:
+4.8 hPa (+4.8 cm H2O)
Demand valve flow spontaneous breathing:  0 -100 lpm
Relief valve triggering pressure: 55 hPa (55 cm H2O)
Alarm valve triggering pressure: 46 hPa (46 cm H2O)

MARS II HOSE ASSEMBLY

Connection Probes: by the national standards
Working pressure: 4.2 bar
Burst pressure: ±56 bar / 23°C and ±40 bar / 40°C
Material:  PVC, anti-static in accordance with ISO 5359
Lenght: 1.2 m
Diameter: from 14 mm

MARS II ACCESSORIES

Art. Nr. Description
8848004   Reusable universal size mask 
1035575  Breathing filter
1032907  Adult mask
1035329  Airway guedel no 3
1035328  Airway guedel no 2
1035481  Airway guedel no 1
1033026  Net head harness
SPP35810044  Cover hose assembly
1024374  Red square Mars bag

MARS II REGULATOR

Pressure gauge (reg): 0-200 bar
Therapy flows (reg): 1 to 25 lpm
Therapy outlet: Universal firtree 6 mm
Outlet pressure: 4.2 bar
Pressure relief valve setting: 8.4 bar
Quick connector: Regional standard according to probe
Standard: ISO 10524-1
Operating temperature: -18°C to +50°C
Storage temperature: -40°C to +60°C
Manufacturer: GCE, s.r.o, Žižkova 381, 583 81 Chotěboř, CZ
CE-marking: CE2460